Converter manometric ionization PMI-2
Also, this product can be called: PMI2, PMI 2, pmi-2, pmi2, pmi 2.
PMI-2 manometric ionization transducer converts the pressure value into an electrical signal, designed to work in conjunction with vacuum gauges or other similar devices.

Specifications PMI-2:
Measurement range - from 10-3 Torr to 10-7 Torr (from 0.13∙10-4 Pa to 0.13 Pa).
The required value of the emission current is 0.5 mA.
The lamp body is C52-I glass.
Accession - "fungal" (choke).
Connecting diameter - 15 mm.
Constant C of products converters PMI-2:
- at an emission current of 5 mA - (1.16±0.23)∙10-3 Pa/A;
- at emission current of 0.5 mA - (11.6±2.3)∙10-3 Pa/A.
Working position - vertical.

Converter resource:

- at a pressure in the vacuum system of 0.13∙10-2 Pa - 750 h;
- at a vacuum pressure of 0.13∙10-1 Pa - 75 h;
- at a vacuum pressure of 0.13 Pa - 7.5 hours.
The pressure in the examined volume to ensure safe operation is not more than 0.13 Pa.
Warranty operating time - 500 hours.
Operation at pressures from 0.13 Pa to 0.013 Pa leads to its premature failure.

Overall dimensions PMI-2:
- diameter - 34 mm;
- height - 265 mm.
The stem diameter is 16.3 mm.

The measured pressure is calculated by the formula P = C/I, where I is the measured ion current, A; C is the converter constant.

The PMI-2 lamp is attached to the vacuum volume either with the help of a fungal (compression) connection, or with the help of a glass-metal junction. For convenient connection of the sensor to the vacuum gauge, it is recommended to use the connecting cable KS-PMI2.
To measure pressure using this sensor, first, the degassing procedure is carried out (cathode and anode heating), then calibration (measurement of "zero" currents) and emission setting (high voltage is connected to the anode and the required emission current value is set by changing the cathode heating current).
If the PMI-2 ionization sensor was turned off, then repeated degassing is not required. If the device is switched off, degassing will be restarted.

Combine shipping is accepted.
If you need less or more items of this type, or if you need add some other items, contact us, and we will make special listing for you.
Give us the list of items you need, and we'll do the rest.

We have many other vacuum tubes for audio and other, ask us if you need any of them.

Items are shipped from Lithuania Post via expedited International tracking.

View full details